SECTION 6—TROUBLESHOOTING
31
SECTION 6—TROUBLESHOOTING
NOTE: If problems are not remedied by the suggested means above, contact an Invacare Dealer.
SYSTEM/OPERATION
SYMPTOMS
FAULTS
SOLUTION
Adjustable Height
Model IH6300ADH
Tub will not raise
or lower
Glide pads not greased
Grease glide pads.
Refer to Maintenance for
IH6300ADH Adjustable Height
ONLY Glide Pads on page 26.
Bad electrical connections
Check electrical connections at
the control box.
Actuator overloaded
Wait 5 minutes and try adjusting
height again.
Actuator burned out
Contact an Invacare Dealer.
Obstacle in tub path
Inspect and remove obstacle.
No electrical power to tub Check and reset breaker or
GFCI.
Hydromassage System
System will not start
No electrical power to tub Check and reset breaker or
GFCI.
System not prepared
Be sure water level is above jets.
Check that inlet plug on tub is
depressed.
Switch hose disconnected
Reconnect hose to switch and
motor.
System is ineffective
Water deposits in
whirlpool jets
Disassemble and clean whirlpool
jets.
Whirlpool jets closed
Open whirlpool jets.
Aerator closed
Open aerator.
Whirlpool intake clogged
Inspect intake and remove
obstruction.
Disinfecting System
No discharge from
disinfectant wand
Whirlpool jets open
Close whirlpool jets.
Whirlpool intake open
Close whirlpool intake.
Hose kinked
Inspect hose and unkink.
Disinfectant siphon tube
kinked
Inspect siphon tube and unkink.
Flow meter adjustment
knob turned off
Turn flow meter adjustment knob
to proper setting.
Tub Operation
No water flow
Water supply turned off
Turn on water supply.
Restricted water flow RADA valve. Refer to
RADA owner’s manual.
Inspect and clean RADA valve.