Calibrating the disinfectant flowmeter, Fore performing this procedure. refer to – Invacare IH6300ADH User Manual

Page 28

Advertising
background image

SECTION 5—MAINTENANCE

Supine Bath Tub

28

Part No 1096926

Calibrating the Disinfectant Flowmeter

ƽ

WARNING

DO NOT use this product or any available optional equipment without first completely reading and

understanding these instructions and any additional instructional material such as owner’s manuals,

service manuals or instruction sheets supplied with this product or optional equipment. If you are

unable to understand the warnings, cautions or instructions, contact a healthcare professional,

dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage

may occur.
The disinfectant dispensing system has been factory tested and calibrated for use with Invacare

disinfectant. The use of any disinfectant with a dilution ratio (½ oz. per gallon) or viscosity level

different than that of Invacare disinfectant will result in an improper mixture of disinfectant and

water. This may cause your system to be improperly disinfected - or in some instances, not

disinfected at all. If using any disinfectant with a dilution ratio or viscosity level different than that of

Invacare disinfectant or if switching from one type of disinfectant to another type, the disinfectant

flowmeter MUST be calibrated using this procedure before disinfecting the tub.
This procedure should only be attempted by qualified maintenance personnel. Failure to heed these

precautions could result in injury to the tub operator or resident.
ALWAYS wear rubber gloves, an apron and face shield protection during disinfecting, cleaning and

rinsing procedures. In case of eye or skin contact, follow procedures on disinfectant container.

NOTE: For this procedure, refer to FIGURE 5.2 on page 29.
NOTE: Equipment needed:

Bucket or Pail, at least 1 Gallon

Graduated plastic bottle, at least 16 ounces

Felt Tip Pen

1.

Open the whirlpool aerator by rotating two turns counterclockwise.

2.

If the tub is equipped with a disinfectant wand, hold the wand over the interior of the tub.

3.

Pour exactly one gallon of water into the bucket/pail.

4.

Mark the bucket/pail at the gallon level using the felt tip pen.

5.

Empty the bucket/pail.

6.

Pour a whole number of ounces of disinfectant into the graduated plastic bottle.

NOTE: For example, pour 2 or 3 ounces, but not 2½ ounces, otherwise the results will be skewed.
NOTE: DO NOT use water instead of disinfectant. This would cause a false reading.
7.

Note the number of ounces in the graduated plastic bottle.

8.

Open the service access cabinet door.

9.

Insert the disinfectant draw tube into the graduated plastic bottle.

10. Turn the black knob on the flow meter valve five turns counterclockwise to open the metering valve.
11. Close all jets by pushing the center of each jet in or turning the center of the jet clockwise.
12. Close the whirlpool inlet by pulling the inlet plug out.
13. Turn the disinfectant valve to the on position.
NOTE: Liquid will be heard filling the interior plumbing of the tub.
14. Note the position of the floating ball in the flow meter. 

Advertising
This manual is related to the following products: