14 TROUBLESHOOTING

Lynx™L-3X Scooter

58

Part No. 1145807

Scooter will not drive.

Error

Brake release lever disengaged.

Batteries require charging.

Key not fully plugged in.

Charger plugged in.

Fuse blown in battery box.

Soft stop engaged. If key is turned off while

driving, the scooter holds power for 1.5 seconds

slowly decelerating to a stop.

Battery connections loose.

Check service indicator light on control panel (FIGURE 6.1 on

page 25). If flashing, count the number of flashes and refer to

Service Indicator Error Codes on page 59.

Engage brake release lever. Refer to Engaging/Disengaging the

Brake Release Lever on page 22.

Charge batteries. Refer to Charging the Batteries on page 32.

Ensure key is inserted fully into keyswitch.

Unplug charger from scooter.

Contact Dealer or qualified technician.

With key in the off position, wait for 30 seconds and turn key

to the on position.

Check all connections. Secure connections. Refer to

Removing/Installing the Battery Box on page 30.

SYMPTOM

PROBABLE CAUSE

SOLUTION