SECTION 11—BATTERIES

Part No 1125085

67

Pronto® M51™and M61™with SureStep®

CAUTION

New batteries MUST be fully charged prior to initial use of the wheelchair.
ALWAYS charge new batteries before initial use or battery life will be reduced.
As a general rule, you should recharge your batteries as frequently as possible to

assure the longest possible life and to minimize required charging time. Plan to

recharge them when you do not anticipate using the wheelchair.

Basic concepts which will help you understand this automatic process are:

The amount of electrical current drawn within a given time to charge a battery is called 
“charge rate”. If, due to usage, the charge stored in the battery is low, the charge rate is 
high. As a charge builds up, the charge rate is reduced, and the battery charger rate 
decreases to a “trickle charge”.
NOTE: If the batteries need to be charged more often or take longer to charge than normal, they 
may need to be replaced. Contact a qualified technician.

NOTE: The batteries can be charged using the on‐board battery charger or by plugging an 
independent battery charger into the port located on the front of the joystick.

Battery Charger Operation

ƽ WARNING

Read and carefully follow the manufacturer’s instructions for each charger (supplied

or purchased). If charging instructions are not supplied, consult a qualified

technician for proper procedures.
NEVER leave the charger unattended when the charger circuit breaker is tripping.

DANGER

Use of improper extension cord could result in risk of fire and electric shock.