Sensor validation procedure – Sierra FLAT-TRAKTM 780S User Manual

Page 41

Advertising
background image

Series 780S Instruction Manual

Chapter 3 Operation

IM-78S-B

3-15

Calibration Certificate Values

Validation Test Results

Sample

Point

Bridge

Voltage

Indicated

Flow

Output

(V or mA)

Indicated

Flow

(LCD)

Flow

Meter

Stated

Accuracy

Output

(V or mA)

Flow

Meter

Stated

Accuracy

0%

25%

50%

75%

100%

Table 3-1. Electronics Validation Results

Sensor Validation Procedure

1. Locate the Ro temperature (measured resistance at 0°C) value and

the Alpha value shown on the Calibration Certificate supplied with
the flow meter.


2. Turn off power to the flow meter. Allow a 6 minute

cool-down before continuing.


3. Remove the cover of the flow meter enclosure to access the sensor

connection points. Remove the four-position jumper from J5, J6,
J7 and J8 (see below for location).

Smart Electronic Device
NEMA 4X Enclosures

Remove jumper, measure
on RIGHT ROW of pins

J5 temperature
J6 temperature
J7 velocity
J8 velocity

1
2
3
4
5
6
7
8
9

10

20
19
18
17
16
15
14
13
12
11

Terminal Board
Hazardous-Area Enclosures

Remove jumper,
measure on LEFT
ROW of pins

J5 temperature
J6 temperature
J7 velocity
J8 velocity

Figure 3-4. Sensor Validation Component Locations

Warning!

Do not power the flow

meter with the sensor
jumper disconnected.

This could cause over-

heating of the sensors

and/or damage to the

electronics.