Appendix a product specifications – Sierra FLAT-TRAKTM 780S User Manual

Page 47

Advertising
background image

Series 780S Instruction Manual

Appendix A Specifications

IM-78S-B

A-1

Appendix A

Product Specifications

Performance Specifications

Accuracy

± 2% of reading from 10 to 100% of calibrated range
± 0.5% of full scale below 10% of calibrated range

Repeatability

± 0.2% of full scale

Temperature Coefficient

± 0.02% of reading per °F within ± 50°F of customer specified conditions
± 0.03% of reading per °F within ± 50°F to 100°F of customer specified conditions
±0.04% of reading per °C within ± 25°C of customer specified conditions
±0.06% of reading per °C within ± 25°C to 50°C of customer specified conditions

Pressure Coefficient

Negligible when within ± 50 psig (± 3.4 barg) of customer specified conditions

Response Time

One second to 63% of final velocity value

Operating Specifications

Gases

Most gases compatible with 316L stainless steel (consult factory)

Mass Flow Rates

Air Flow Ranges

(1)

Pipe Size

Minimum

Maximum

(2) (3)

scfm

(nm

3

/hr)

scfm

(nm

3

/hr)

1/4-inch

0–0.5

(0–0.7)

0–9

(0–14)

1/2-inch

0–2

(0–3.0)

0–40

(0–60)

3/4-inch

0–4

(0–5.9)

0–75

(0–120)

1-inch

0–6

(0–8.9)

0–120

(0–180)

1 1/2-inch

0–15

(0–22)

0–280

(0–440)

2-inch

0–23

(0–33)

0–470

(0–680)

3-inch

0–50

(0–74)

0–1000

(0–1500)

4-inch

0–90

(0–130)

0–1800

(0–2700)

6-inch

0–200

(0–300)

0–4000

(0–5900)

8-inch

0–350

(0–520)

0–7000

(0–10,000)

Notes:

(1) Flow rates for air and nitrogen. Standard conditions: 70°F
and one atmosphere for scfm; 0°C and one atmosphere for
nm

3

/hr. Consult factory for other gases.

(2) Higher flow ranges available, consult factory.
(3) Maximum flow rates are limited for hazardous area
and high temperature versions, consult factory.

Dual Calibration

User-selectable dual ranges or two different gases

Gas Pressure

150 lb flange or PN16 DIN (–40° F to 100° F): 230 psig (15.9 barg) maximum
150 lb flange or PN16 DIN (250° F): 185 psig (12.8 barg) maximum
150 lb flange or PN16 DIN (450° F): 155 psig (10.7 barg) maximum
NPT (–40° F to 450° F): 500 psig (34 barg) maximum, 5 to 150 psig
(0.3 barg to 10 barg) optimum

Gas & Ambient Temperature Gas...............–40° to 250°F (–40° to 120°C) Optional –40° F to 450° F

(–40° C to 230° C) not available on 1/4 and 1/2-inch flow bodies
Ambient........–5° to 120°F (–20° to 50°C)