Chapter 4 Troubleshooting & Repair

Series 780S Instruction Manual

4-2

IM-78S-B

Problem

Possible Cause

Solution

Velocity measurement is
erratic or fluctuating

Very erratic or non-uniform flow

Follow installation requirements shown
in Chapter 2

Moisture present in gas flow

Install a water trap or filter upstream of
the flow meter sensor

Flow conditioning plates are not
upstream of the sensor

Correct flow meter orientation

Sensor component broken

Return to factory for replacement

Malfunction in system
electronics

Return to factory for evaluation

Ground loop

Check wiring, see Chapter 2

Velocity measurement
seems too high or low

Sensor assembly not aligned
correctly to flow

Correct alignment with the flow indicator
pointing downstream in the direction of
flow

Flow conditioning plates are not
upstream of the sensor

Correct flow meter orientation

No response to flow from
sensor assembly

No power

Turn on power to the flow meter

Low flow cutoff setting too high

Correct low flow cutoff programming
using the Smart Interface software.

Flow rate below meter’s minimum
flow rating

Contact factory for instructions

Flow has exceeded the maximum
range of the flow meter

Set the user full scale to equal the fac-
tory full scale

Reduce flow below the maximum range
shown on the meter’s nameplate or con-
tact the factory for re-calibration advice

Sensor failure

Return to factory for evaluation

Printed circuit assembly defective

Return to factory for evaluation