Lakeshore Learning Materials 622 User Manual

Page 13

Advertising
background image

Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Introduction

1-3

Front Panel: Contains a menu driven keyboard and graphic display for entry and display of results. Operating

parameters to set and monitor from the front panel (and remote interface) include:

• Output current and compliance voltage setting.
• Output current and voltage measurement.
• Status reporting.
• Output ramp programming.
• Persistent switch heater control (with optional Model 6228 Card).
• Liquid helium level monitoring (with optional Model 6226/6476 Card).
• Field Monitoring (with optional Model 6226 Card).
• Output Current Zeroing.
• Output Current Step Limiting.

Magnet inductance and compliance (di/dt = V

SET

/L) limit the output ramp programming charging current.

Program output for a constant 0.01 to 99.99 amperes per second as long as compliance is not exceeded.

Energize or de-energize the magnet at a pre-set ramp rate. Pause the ramp at any time during the ramp.

During a pause, the MPS maintains output values until the ramp continues.

Operating Ambient Temperature: 15 to 35 °C (59 to 95 °F)

Dimensions: 483 mm wide x 178 mm high x 508 mm deep (19 x 7 x 20 inches)

Weight: 43.5 kilograms (96 pounds). Rack mounting is standard.

Table 1-3. Model 620/622/623 DC Output Specifications

SPECIFICATION CURRENT

VOLTAGE

Digital Programming Resolution for 620/622

1 mA

1 mV

Digital Programming Resolution for 623

1.2 mA

1 mV

Digital Programming Accuracy

0.1% I

MAX

±30 mV

Digital Programming Repeatability

0.01% I

MAX

±30 mV

Electronic Resolution for 620/622

1 mA

1 mV

Electronic Resolution for 623

1.2 mA

1 mV

Electronic Accuracy

0.1% I

MAX

±30 mV

Display Resolution

1 mA

1 mV

Stability (Drift) at 25 ±1 °C:

Percent of full scale

output change over 8-hours under constant line and load after
a 30 minute warm-up.

±0.005% I

MAX

±3 mV

Ripple and Noise:

10 Hz to 10 MHz at 1000 VA

20 µA rms

10 mV rms

Temperature Coefficient:

Change in output per °C

after 30 minute warm-up.

±0.01% I

MAX

±3 mV

Source Effect:

Line regulation for any line change within

the rated line voltage.

±0.005% I

MAX

±15 mV

Load Effect:

Load regulation for a load change equal to

maximum voltage in Constant Current Mode or maximum
current in Constant Voltage Mode.

±0.01% I

MAX

±3 mV

Analog Resistance Programming Accuracy:

0 to 10 K

produces negative full scale to positive full scale

current or voltage output. 5 K

is 0 current.

10% I

MAX

3 V

Analog Voltage Programming Accuracy:

Voltage input is ±0.01 V/A, ±0.01 V/V.

1% + 100 mA

2% + 100 mV

Monitoring Output Accuracy:

Voltage output is

±0.01 V/A, ±0.01 V/V.

1% + 100 mA

2% + 100 mV

Advertising
This manual is related to the following products:

623, 620