Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Operation

3-10

2. Enter these parameters in the Automatic Mode Screen:

MAG START: Magnet Start Current (only if
LAST END is unknown). For this example,
we assume an initial MPS hook-up to this
magnet and a previous magnet charge to
20 A.

MGT

END: Magnet End Current. Enter 30

A.

MGT

RATE: Magnet Charging Rate. Base this figure on the maximum magnet current supplied by the

magnet manufacturer. For this example, enter 0.4 A/S (amperes per second).

PERS

RATE: Persistence Switch Charging Rate. Base this figure on the maximum Persistence Switch

current (generally much higher than the magnet charging rate). Anything over approximately 20 A/S

appears instantaneous. For this example, enter 4 A/S.

3. Enter the following parameters in the Automatic Delay Screen (Function Menu 3).

RUN UP: The time between initial ramp up
to current and PSH activation (Persistence
Mode OFF). Enter 5 seconds.
PERS OFF: The time between PSH

activation (Persistence Mode OFF) and

initial ramp to setpoint. Enter 5 seconds.
RAMP END: The time between ramp up to setpoint and PSH deactivation (Persistence Mode ON). Enter

5 seconds.
PERS ON: The time between PSH deactivation (Persistence Mode ON) and initial ramp back to zero.

Enter 5 seconds.

4. Return to the Automatic

Mode Screen. Move to the

RUN line, then toggle to

activate Automatic Mode.

The MPS automatically runs

through the cycle as defined,

providing messages on the

bottom line of the display to

detail the current step of the

cycle. When complete, the

display returns to

STANDBY.

After determining proper system

Delay times, it is unnecessary to

visit the Delay Screen each time;

the delays store in non-volatile

memory. After the MPS

completes the cycle once, it

stores the ending current value

to LAST END. To make

subsequent changes to magnet

current simply enter a new MGT

END setpoint.

To pause a ramp or skip a time out, press either the Data Entry up or down key. To abort the operation, press

Escape. After an abort, <RUN> continues an automatic process from where it left off.

A U T O M A T I C M O D E :

¤ M G T S T A R T

M G T E N D

M G T R A T E

P E R S R A T E

L A S T E N D

P E R S I S T E N C E

R U N ¥

O U T P U T S :

*

+0.000A

: 0 . 0 0 0 A

: 0 . 0 0 0 A

: 1 . 0 0 0 A / S

: 1 . 0 0 0 A / S

: 0 . 0 0 0 A

: < O F F >

+ 0 . 0 0 1 V

S T A N D B Y

A U T O M A T I C M O D E D E L A Y S :

¤ R U N U P

P E R S O F F

R A M P E N D

P E R S O N

O U T P U T S :

*

+1.999A

: 6 0 S

: 6 0 S

: 6 0 S

: 6 0 S

L V L : + 1 4 . 4 i n

F L D : + 6 3 . 4 8 k G

+ 0 . 0 0 1 V

0

ON

OFF

5

PERS RATE Ramp
Delay
MGT RATE Ramp

RAMP

UP

Delay

PERS

OFF

Delay

RAMP

END

Delay

PERS

ON

Delay

MGT

START

Persistence Off

(PSH On)

MGT

END

Persistence On

(PSH Off)

LAST

END

AUTOMATIC MODE

5

10

15

20

25

30

35

MPS Current Output (amperes)

P

e

rs

istence

Mode Status

10

15

20

25

30

35

40

45

50

57.5

Time (seconds)

Figure 3-2 Typical Automatic Mode Cycle