Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Error Messages & Troubleshooting

5-3

High AC (HI AC) Detected

Any time the AC line rises above 120% of the

nominal line selected, the MPS turns off the front

panel circuit breaker and enters AC Loss Mode.

This halts excessive utility voltages to internal

circuitry. If this occurs on power up, the MPS turns

off the front panel circuit breaker and displays:

+ 0 . 0 0 1 V

* * H I A C L I N E * *

O U T P U T S :

*

+1.999A

T H E U N I T W I L L B E T U R N E D

O F F T O A V O I D D A M A G E .

V E R I F Y A C I N P U T S E T T I N G

HI AC INPUT

LINE FAULT:

Overtemperature Detected
If the MPS detects internal over temperature or

other internal fault, it activates the OVP circuit,

turns off the front panel circuit breaker, and then

turns itself off. OVP stays active until it completely

discharges the load. If the heat sink temperature

exceeds 95 °C (203 °F), the MPS displays:

+ 0 . 0 0 1 V

* * T E M P F A U L T * *

O U T P U T S :

*

+1.999A

T H E U N I T W I L L B E T U R N E D

O F F T O A V O I D D A M A G E .

I N T E R N A L T E M P T O O H I

INTERNAL

TEMP FAULT:

Rail Fault Detected

Excessive DC rail voltage (unable to return power

to the utility or some other internal fault), causes

the MPS to display:

+ 0 . 0 0 1 V

* * R A I L F A U L T * *

O U T P U T S :

*

+1.999A

T H E U N I T W I L L B E T U R N E D

O F F T O A V O I D D A M A G E .

D C R A I L V O L T A G E T O O H I

INTERNAL

RAIL FAULT:

PSH Current Not The Same As Saved Value
When the PSH is turned ON, the MPS checks the

present current setting against the current setting

stored when the PSH was turned OFF. If the two

do not match, the MPS sounds an alarm for about

½ second and displays:

C A U T I O N ! S E T T I N G N O T S A M E A S S A V E D V A L U E

L A S T P E R S I S T E N C E C U R R E N T :

P R E S E N T S U P P L Y C U R R E N T :

E N T E R K E Y T O C H A N G E T O L A S T P E R S I S T E N C E

C U R R E N T , W A I T R U N U P D E L A Y , T H E N T U R N O F F .

¥ K E Y T O T U R N O F F A T P R E S E N T C U R R E N T .

E s c K E Y T O A B O R T .

H E A T E R C U R R E N T :

5 0 m A

1 0 . 0 0 0 A

5 . 0 0 0 A

Press the Data Entry up or down arrow to turn the
PSH ON. Press no key for approximately 30
seconds, or press Esc to keep the PSH OFF.

Output Inhibit
When the red Output Inhibit button is pressed, the
MPS displays:

+ 0 . 0 0 1 V

* * O U T P U T I N H . * *

O U T P U T I N H I B I T :

T H E O U T P U T C U R R E N T S E T -

T I N G I S F O R C E D T O Z E R O .

T H E O U T P U T S E T T I N G S

C A N N O T B E C H A N G E D U N T I L

O I I S R E L E A S E D .

O U T P U T S :

*

+1.999A

+ 0 . 0 0 1 V

* * O U T P U T I N H . * *

O U T P U T S :

*

+1.999A

I STEP Limit Exceeded
When the user attempts to change the current setting
by more than the step limit setting, the MPS displays:

O U T P U T C U R R E N T W O U L D B E S E T T O M O R E T H A N

I S T E P L I M I T P E R M I T T E D I N 1 U P D A T E .

T H E C U R R E N T S E T T I N G W I L L N O T B E C H A N G E D .

I STEP LIMIT EXCEEDED

BY THIS SETTING

Caution – Persistence Is Off, Heater Is On
When the user attempts to start Automatic Mode with
persistence Off, the MPS displays:

T H E P E R S I S T E N C E I S O F F .

T H E S W I T C H H E A T E R I S O N .

C A N N O T B E G I N W I T H H E A T E R O N .

H I T T H E E s c K E Y T O A B O R T .

CAUTION

Caution – Same Magnet Start and End Current
To enter an automatic PSH cycle, the magnet start and
end current must be different. If they are the same, the
MPS displays:

T H E M A G N E T S T A R T C U R R E N T A N D T H E M A G N E T

E N D C U R R E N T A R E T H E S A M E . P R E S S T H E E s c

K E Y T O A B O R T , T H E E N T E R K E Y T O C O N T I N U E .

CAUTION

Maximum Power Output Exceeded
When users utilize external current programming and
output power exceeds 1000 watts, the MPS displays:

+ 0 . 0 0 1 V

* * O V E R P O W E R * *

M A X I M U M P O W E R O U T P U T

E X C E E D E D :

T H E O U T P U T P O W E R I S

G R E A T E R T H A N P E R M I T T E D .

S E E M A N U A L F O R L I M I T .

T H E O U T P U T C U R R E N T I S

F O R C E D T O Z E R O A M P E R E .

P R E S S ¥ T O C L E A R E R R O R .

O U T P U T S :

*

+1.999A

O U T P U T S :

*

+1.999A

Front panel settings are software limited to prevent
power over 1000 watts.

Advertising
This manual is related to the following products:

623, 620