Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Remote Operation

4-19

*IDN?

Identification Query

Input:

*IDN?

Returned: Manufacturer, Model Number, Model Serial Number, Firmware Date.

Remarks: Returns seventeen characters plus up to two terminators.

Example: LSCI,622,0,120193[term]

*OPC

Operation Complete Status

Input:

*OPC

Returned: Nothing.

Remarks: Causes the MPS to set an update cycle counter to 2. Each time through an update cycle,

parameters entered by device dependent commands update and the cycle count decrements.
Two update cycles complete all pending device dependent commands. When this occurs, the
MPS sets the operation complete bit in the Standard Event Status Register (not the
IEEE-488.2 defined operation). Not supported by Control Bus.

*OPC?

Operation Complete Status Query

Input:

*OPC?

Returned: 0 or 1, where 0 indicates incomplete, 1 indicates complete.

Remarks: Places 1 in the instrument output queue and sets the Operation Complete Bit in the Standard

Event Status Register upon completion of all pending selected device operations. This must be
sent as the last command in a command string. Returns one character plus up to two
terminators. Not supported by Control Bus.

*RST

Instrument Reset Command

Input:

*RST

Returned: Nothing.

Remarks: Restores MPS to power-up settings just like the DCL and SDC bus commands. The MPS

reverts to the Normal Display Screen, forces ramp status to the Hold mode, and clears any
latched error along with the Status Byte and the Standard Event Status Registers. Not
supported by Control Bus.