Chapter 3, Operation, Figure 3-1 mps front panel – Lakeshore Learning Materials 622 User Manual

Page 27

Advertising
background image

Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Operation

3-1

CHAPTER 3

OPERATION

3.0 GENERAL
This chapter covers seven areas: MPS Front Panel (Paragraph 3.1), Power Up (Paragraph 3.2), Setting

Current on an MPS with Manual PSH Control (Paragraph 3.3), Setting Current with Automatic PSH Control

(Paragraph 3.4), Instrument Setup Screens (Paragraph 3.5), Function Menus (Paragraph 3.6), and an

Automatic Persistence Control Example (Paragraph 3.7).

Setup the MPS software with the MPS output terminals shorted together as in section 3.2. This ensures that

while the user learns MPS operation, an inadvertent keystroke causes no damage to the magnet.

3.1 THE MPS FRONT PANEL
Figure 3.1 below shows the MPS Front Panel. The up or down Display Cursor Control keys move the line

indicator to the line containing a value to be changed. Use either the Data Entry Keypad to enter the desired

value or the up or down Numeric Entry keys to increment or decrement the value. The Enter key accepts the

update, while the Esc key discards the change and returns to the prior value. Move the cursor to a particular

digit with the right and left Display Cursor Control keys and change it with the Numeric Entry keys.

7

8

9

4

5

6

1

2

3

0

.

+/-

Enter

Esc

Function

Menu

Normal
Display

Next

Menu

Cursor

Local

Remote

Mode

Output

Inhibit

Functions

Power

Fault

Persistent

Switch Heater

OFF

ON

Magnet Power Supply

Data Entry

A

B

C

D

I

E

F

H

N

K

L

M

J

G

A. Display Screen

B. Display Function Keys

C. Display Menu Selection Keys

D. Data Entry Keypad

E. Escape Key. Restores an old value and

returns from a Function Screen to

Normal Screen. Also Aborts a ramp or

automatic process.

F. Interface Keys

G. Data Entry Up and Down Keys. Increment and

decrement numeric values, Pause/Restart

ramps, or execute some functions.

H. Output Inhibit Key
I. Display Cursor Control Keys
J. Enter Key. Accepts a change.
K.
Power Indicator
L. Fault Indicator
M. Persistent Switch Heater Indicator
N. Power Switch

Figure 3-1 MPS Front Panel