Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Mainframe Control Bus Operation

B-2

B4.0 CONTROL BUS SERIAL INTERFACE CONFIGURATION
Configure the control bus serial interface for either RS-232C or RS-485 communications with DIP switches on

the Communications Processor Board (CMP). The CMP plugs into the Mainframe rear panel. The standard

control bus serial interface configuration is RS-485. Use the procedure below to change or verify the serial

interface configuration.

1. Turn off Mainframe and unplug the power cord.

2. Loosen the two screws that secure the CMP to

the rear panel and slide the CMP out.

3. Locate and configure the control bus serial

interface DIP switches as defined in the table to

the right. The DIP switches are just inside the

CMP front panel. Switches 1-4 define the

interface as RS-232C. Switches 5-8 define the

interface as RS-485 multidrop. Switch 8

activates 100

bus termination for long serial

interface runs. For RS-485, terminate extremely

long main lines at the furthest point from the

host. Limit main line length to under 4000 feet

and multiple short stub lines from the main line

(Mainframe to Mainframe) to under 50 feet. A typical RS-485 configuration runs the main line from the

host to one of the first Mainframe serial interface jacks; a Model 2001 cable plugs into the other first

Mainframe serial interface jack to connect to the second Mainframe, and so on. Connect up to 32

Mainframes in this way. Terminate the last Mainframe only if it is a significant distance from the host.

Assign unique addresses to multiple Mainframes on a RS-485 interface.

4. Replace the CMP.

B5.0 SELECTING THE CONTROL BUS SERIAL INTERFACE ADDRESS
Multiple Mainframes on the RS-485 interface each have a unique address and operate in a polled mode. Set

the MPS Mainframe to respond to one of 32 addresses. Prefix all commands sent to a Mainframe with the

Mainframe address. A Mainframe query response includes the address prefix if the address is not 0. The

receipt of the LF (0AH) terminator resynchronizes communications and alerts all Mainframes that the next

character is an address. A Mainframe with address 0 responds to all commands on the interface. A Mainframe

using RS-232C can be given a unique address as well, but RS-232C is not suited for multi-drop operation.

Access the Control Bus serial interface address from the Interface Setup screen. Press the Function Menu

key to display Function Menu 1. Press the key corresponding to INTERFACE SETUP to display the Interface

Setup screen.

Use the cursor key to move to the line indicating CB DEF and CB ADD; defined as follows.

CB DEF: 9600 Baud, No Parity, 8 Data Bits, and 1
Stop Bit

Control Bus Definition. The control bus should be

defined as above.

CB ADD: 0 to 32

Control Bus Address. Use the Data Entry up or

down arrows to increment or decrement the Control Bus Address. The initial condition is 0.

CMP

Configuration

DIP Switch

RS-232C Terminated

RS-485

Unterminated

RS-485

1 Closed

Open Open

2 Closed

Open Open

3 Closed

Open Open

4 Closed

Open Open

5 Open

Closed

Closed

6 Open

Closed

Closed

7 Open

Closed

Closed

8 Open

Closed

Open

I N T E R F A C E S E T U P :

¤ I E E E A D D

T R M S / E O I

R S 2 3 2 C

C B D E F

C B A D D

V W A N G L E

M F I D

O U T P U T S :

*

+1.999A

: 1 2

: (CR LF) EOI

: 9 6 0 0 , N , 8 , 1

: 9 6 0 0 , 0 , 7 , 1

: 0

: 9

: 6 2 2 ( 1 )

¥

¥

¥

¥

¥

+ 0 . 0 0 1 V