Common Errors and Their Solutions

Appendix B

GPIB-PC User Manual

B-6

©National Instruments Corp.

EBUS(14)

Error Condition: Command error during device call.

Solutions:

Find out which device is abnormally slow to accept commands
and fix the problem with the device.

If more time is needed to send commands, lengthen the time
limit of the board in

IBCONF

or with

IBTMO

.

ESTB(15)

Error Condition: Serial poll status byte(s) lost.

Solutions:

Call

IBRSP

more often to read the status bytes.

Ignore ESTB.

ESRQ(16)

Error Condition: SRQ stuck in the ON position.

Solutions:

Ignore ESRQ until all devices are found. It occurred because
the device asserting SRQ was not opened with

IBFIND

. The

automatic serial polling polls only the opened devices.

Check that you have used

IBFIND

to open all devices on the

GPIB that could assert SRQ. Remove any device from the bus
if it is not being accessed.

Using

IBIC

, attach one device at a time and determine that it

is unasserting SRQ after being polled.

Inspect the interconnecting cable to see that the devices are
attached and that the connectors are seated properly.