Analog input timing signal, Ai sample clock, Figure 11. sample clock timing options – National Instruments NI WLS/ENET-9163 User Manual

Page 21: Convert behavior for analog input modules, Scanned, Analog input timing signal ai sample clock, Convert behavior for analog input modules scanned

Advertising
background image

© National Instruments Corporation

21

NI WLS/ENET-9163 Carrier User Guide and Specifications

Note

Pause triggers are only sensitive to the level of the source, not the edge.

Analog Input Timing Signal

AI Sample Clock

A sample consists of one reading from each channel in the AI task.
ai/SampleClock signals the start of a sample of all analog input channels
in the task. ai/SampleClock can be generated from external or internal
sources.

Using A Digital Source

To use ai/SampleClock, specify a source and an edge on PFI 1. For more
information, refer to the Device Routing in MAX topic in the NI-DAQmx
Help
.

Figure 11. Sample Clock Timing Options

Routing AI Sample Clock to an Output Terminal

You can route ai/SampleClock to the PFI 1 terminal.

Convert Behavior For Analog Input Modules

Scanned

Scanned C Series analog input modules contain a single A/D converter and
a multiplexer to select between multiple input channels. When the C Series
Module Interface receives a Sample Clock pulse, it begins generating a
Convert Clock for each scanned module in the current task. Each Convert
Clock signals the acquisition of a single channel from that module. The
Convert Clock rate depends on the module being used, the number of
channels, and the Sample Clock rate.

Programmable

Clock

Divider

ai/SampleClock Timebase

20 MHz Timebase

PFI

ai/SampleClock

Sigma-Delta Module Internal Output

Advertising