Ni 9472 with dsub safety voltages – National Instruments NI 9472 User Manual

Page 22

Advertising
background image

NI 9472/9474

22

ni.com

This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S.
or 230 V for Europe.

Caution

Do not connect the NI 9472/9474 with screw

terminal to signals or use for measurements within
Measurement Categories III or IV.

NI 9472 with DSUB Safety Voltages

Connect only voltages that are within the following limits.

Channel-to-COM .............................. 30 VDC max

Isolation

Channel-to-channel .................... None
Channel-to-earth ground

Continuous ........................... 60 VDC,

Measurement Category I

Withstand ............................. 1,000 V

rms

, verified by a 5 s

dielectric withstand test

Measurement Category I is for measurements performed on
circuits not directly connected to the electrical distribution system
referred to as MAINS voltage. MAINS is a hazardous live electrical