Shock and vibration, Electromagnetic compatibility – National Instruments NI 9472 User Manual

Page 26

Advertising
background image

NI 9472/9474

26

ni.com

Shock and Vibration

To meet these specifications, you must panel mount the system.
If you are using the NI 9472/9474 with screw terminal, you also
must either affix ferrules to the ends of the terminal wires or use the
NI 9932 backshell kit to protect the connections.

Operating vibration

Random (IEC 60068-2-64)......... 5 g

rms

, 10 to 500 Hz

Sinusoidal (IEC 60068-2-6) ....... 5 g, 10 to 500 Hz

Operating shock
(IEC 60068-2-27).............................. 30 g, 11 ms half sine,

50 g, 3 ms half sine,
18 shocks at 6 orientations

Electromagnetic Compatibility

This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement,
control, and laboratory use:

EN 61326 EMC requirements; Industrial Immunity

EN 55011 Emissions; Group 1, Class A

CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A

Advertising