Figure b-1. ai sample clock and ai convert clock – National Instruments Data Acquisition Device NI USB-621x User Manual

Page 155

Advertising
background image

Appendix B

Troubleshooting

NI USB-621x User Manual

B-2

ni.com

are various methods of achieving this reference while maintaining a high
common-mode rejection ratio (CMRR). These methods are outlined in the

Connecting Analog Input Signals on USB-6210/6211/6212 Devices

and

Connecting Analog Input Signals on USB-6215/6216/6218 Devices

sections of Chapter 4,

Analog Input

.

AI GND is an AI common signal that routes directly to the ground
connection point on the devices. You can use this signal if you need a
general analog ground connection point to the device. Refer to the

When to

Use Differential Connections with Ground-Referenced Signal Sources

section of Chapter 4,

Analog Input

, for more information.

How can I use the AI Sample Clock and AI Convert Clock signals on a
USB-621x
device to sample the AI channel(s)?

USB-621x devices use AI Sample Clock and AI Convert Clock to perform
interval sampling. As Figure B-1 shows, AI Sample Clock controls the
sample period, which is determined by the following equation:

1/sample period = sample rate

Figure B-1. AI Sample Clock and AI Convert Clock

AI Convert Clock controls the convert period, which is determined by the
following equation:

1/convert period = convert rate

This method allows multiple channels to be sampled relatively quickly in
relationship to the overall sample rate, providing a nearly simultaneous
effect with a fixed delay between channels.

Channel 0

Channel 1

Convert Period

Sample Period

Advertising