Chapter 4

Analog Input

NI USB-621x User Manual

4-34

ni.com

Figure 4-22. Differential Connections for Ground-Referenced Signal Sources

With this type of connection, the NI-PGIA rejects both the common-mode
noise in the signal and the ground potential difference between the signal
source and the device ground, shown as V

cm

in the figure.

AI+ and AI– must both remain within ±11 V of AI GND.

Using Non-Referenced Single-Ended (NRSE)
Connections for Ground-Referenced Signal Sources

Figure 4-23 shows how to connect ground-reference signal sources to the
USB-6210/6211/6212 device in NRSE mode.

USB-6210/6211/6212 Configured in DIFF Mode

PGIA

+

+

+

+

V

cm

V

s

Ground-

Referenced

Signal

Source

Common-

Mode

Noise and

Ground

Potential

AI GND

AI SENSE

Input Multiplexers

V

m

Measured

Voltage

Instrumentation

Amplifier

AI+

AI–

I/O Connector