Chapter 4
Analog Input
4-34
ni.com
Figure 4-22. Differential Connections for Ground-Referenced Signal Sources
With this type of connection, the NI-PGIA rejects both the common-mode
noise in the signal and the ground potential difference between the signal
source and the device ground, shown as V
cm
in the figure.
AI+ and AI– must both remain within ±11 V of AI GND.
Using Non-Referenced Single-Ended (NRSE)
Connections for Ground-Referenced Signal Sources
Figure 4-23 shows how to connect ground-reference signal sources to the
USB-6210/6211/6212 device in NRSE mode.
USB-6210/6211/6212 Configured in DIFF Mode
PGIA
–
+
–
+
–
+
–
+
V
cm
V
s
Ground-
Referenced
Signal
Source
Common-
Mode
Noise and
Ground
Potential
AI GND
AI SENSE
Input Multiplexers
V
m
Measured
Voltage
Instrumentation
Amplifier
AI+
AI–
I/O Connector