Chapter 4

Analog Input

NI USB-621x User Manual

4-24

ni.com

Table 4-2. USB-6210/6211/6212 Analog Input Configuration

AI Ground-Reference

Setting

*

Floating Signal Sources

(Not Connected to

Building Ground)

Ground-Referenced

Signal Sources

Examples:

• Ungrounded thermocouples

• Signal conditioning with

isolated outputs

• Battery devices

Example:

• Plug-in instruments with

non-isolated outputs

Differential (DIFF)

Non-Referenced
Single-Ended (NRSE)

Referenced Single-Ended
(RSE)

*

Refer to the

Analog Input Ground-Reference Settings

section for descriptions of the RSE, NRSE, and DIFF modes and

software considerations.

Refer to the

Connecting Ground-Referenced Signal Sources

section for more information.

+

+

AI+

AI–

AI GND

Signal Source

DAQ Device

+

+

AI+

AI–

AI GND

Signal Source

DAQ Device

Vcm

+

+

AI

AI SENSE

AI GND

Signal Source

DAQ Device

+

+

AI

AI SENSE

AI GND

Signal Source

DAQ Device

Vcm

+

+

AI

AI GND

Signal Source

DAQ Device

Ground-loop potential (V

A

– V

B

) are added

to measured signal.

NOT RECOMMENDED for the

USB-6210/6211/6212

+

+

AI

AI GND

V

B

V

A

DAQ Device

Signal Source