Analog input timing signals, Figure 4-4. analog input timing options, Analog input timing signals -11 – National Instruments Data Acquisition Device NI USB-621x User Manual

Page 44

Advertising
background image

Chapter 4

Analog Input

© National Instruments Corporation

4-11

NI USB-621x User Manual

Analog Input Timing Signals

In order to provide all of the timing functionality described throughout this
section, USB-621x devices have a flexible timing engine. Figure 4-4
summarizes all of the timing options provided by the analog input timing
engine.

Figure 4-4. Analog Input Timing Options

USB-621x devices use AI Sample Clock (ai/SampleClock) and AI Convert
Clock (ai/ConvertClock) to perform interval sampling. As Figure 4-5
shows, AI Sample Clock controls the sample period, which is determined
by the following equation:

1/Sample Period = Sample Rate

20 MHz Timebase

100 kHz Timebase

Programmable

Clock

Divider

Programmable

Clock

Divider

AI Sample

Clock

Timebase

AI Convert

Clock

Timebase

Analog Comparison Event

Ctr n Internal Output

SW Pulse

Ctr n Internal Output

AI Convert Clock

AI Sample Clock

Advertising