Warning, 5 remote control wiring (option) – Omega RD260A User Manual

Page 21

Advertising
background image

2 - 9

2. INSTALLATION

2.2.5 Remote Control Wiring (option)

!

NOTE

The following settings are available only for the model with remote options (R1).

(1) Turn the power switch off.
(2) Remove the transparent cover of the optional terminals.
(3) Connect the wires for remote control to the optional terminals. The “C” is the

terminal common to other 1 to 5.

(4) Cover the terminals with the transparent cover and secure the screws.

Remote control terminals are arranged as follows:

REMOTE

!

1

3

2

4

C

5

Remote Control Terminal Arrangements

Input characteristics:

Input signal

:

Dry contact, open-collector (TTL or transistor) signal

Control functions and signal types:

• Recording start/stop

level

• Chart speed change

level

• Printout messages

trigger more than 250 msec

Signal characteristics :

ON voltage

0.5 V or less (30 mA DC)

Leakage current in OFF state

0.25 mA or less

Signal duration

250 msec or more

Input types

:

Photocoupler isolation (one line common)
Internal isolated power source (5 V±5%)

Dielectric strength

:

500 V DC for one minute between input terminals and
grounding terminal.

To prevent an electric shock, ensure the main power switch is turned off before
wiring.

NOTE

• Use shielded cables for remote control signals to prevent noise pickup. The

shield should be grounded at the recorder’s terminal.

• To prevent an emission of electromagnetic disturbances, separate the remote

control wires from the power supply and input wires at least 0.1 m. Over 0.5
m is recommended.

WARNING

Advertising