6 - 15

6. SET MODE

6.5 Setting Chart Speed (CHART)

Chart Speed (unit: mm/h)

The data for a dot-printing model (10 mm/h to 1500 mm/h) are shown within the bold border.

10

90

375

3000

15

100

450

3600

20

120

600

4500

25

150

720

4800

30

160

750

5400

40

180

900

6000

50

200

1200

7200

60

240

1500

9000

75

300

1800

10800

80

360

2400

12000

Key operation

Indication

Description

Press the [DISP] key for three seconds to enter the SET mode,
and select [

] using the [

] key. Then press the

[ENT] key.

Using the [

] key, select one of the two kinds of chart speed,

the No.1 and No.2 chart speeds. Then press the [ENT] key.

NOTE

The No.2 chart speed identifies the chart speed which takes effect in case the
speed is changed by remote control (optional function).
For models without the remote control function, the setting of the No.2 chart
speed is invalid.

Select a numeric code referring to the table above using the
[

] key and press the [ENT] key.

NOTE

Digital printing will not be carried out when the chart speeds are:

120 mm/h or more for a dot-printing model

When a pen model uses a fast chart speed such as 600 mm/h to 1500 mm/h,
the position error of the time tick of a periodic printout may become as large as
2 cm for 1500 mm/h.

[

] is dispalyed and the setting is completed.

Press the [ENT] key to set another channel; or press the
[DISP] key for three seconds to return to the OPERATION
mode.

NOTE

If you press the [ESC] key in the middle of setting, the indication returns to the

[

] and already set data will be aborted.

6.5 Setting Chart Speed (CHART)

DISP

ESC

PRINT

PRINT

DISP

ESC

PRINT

3 sec

3 sec

CH UP

CH UP

CH UP

CH UP