background image

4 - 13

4. DAILY OPERATION/MAINTENANCE

Pen model

To maintain proper recording functions, it is recommended to clean the plotter-carriage
shaft once a year.
Cleaning procedure is shown below:

Wipe the two shafts, the upside and downside ones for the plotter carriage, with a
piece of soft cloth or paper without fluffing.
If stains are persistent, use several drops of ethyl alcohol with the cloth or paper.

• Do not damage the plotter’s flexible-printed circuit when cleaning the shafts.
• The shafts do not require lubricating oil.

Plastic plate

Flexible board

Plotter carriage
shafts

Dot-printing model

To maintain proper recording functions, it is recommended to lubricate the printer-
carriage shaft once a year.
Lubricating procedure is shown below:

(1) Wipe up dust or oil from the shaft of the printer carriage with a piece of soft cloth

or paper without fluffing.

(2) Take out the tube of lubricating oil from among the spare parts.
(3) Coat the shaft with a few small drops of lubricating oil, but remove excess oil with

a piece of soft cloth or paper without fluffing.

Shaft

4.8.5 Calibration