6. SET MODE

Difference Computation Setting (DELT Setting)

To obtain the difference between the measured values of two channels, proceed as
follows.

Key Operation

Indication

Description

Press the [DISP] key for three seconds to enter the SET mode.
Select the [

] display by using the [

] key.

Then press the [ENT] key.

Use the [

] key to select the desired channel.

Note that channel No.1 cannot be of the ‘DELT’ type.
Press the [ENT] key.

Use the [

] key to select [

].

Press the [ENT] key.

The display which appears will allow you to enter a reference
channel number. Note that the reference channel number must
be lower than that of the channel being set. Therefore channel
No.1 cannot be of the [

] type.

Use the [

] key to select the desired reference channel number.

Press the [ENT] key.

In case of VOLT the display which appears will show the same
minimum value of the recording span as of the reference
channel.
Use the [

] and [

] keys to select the minimum value

(SPAN L).
Press the [ENT] key.
Note that in case of VOLT, the input range of the DELT-channel
must be within the range of the reference channel. In case of TC,
RTD, the minimum/maximum span values can be found by
taking the ±value of the input range width. E.g. input range of
TC type L = –200 to 900°C: minimum/maximum span values
will be at ±1100°C. The difference in measured value from the
reference channel will be recorded on the DELT-channel.

The display which appears will show the same maximum value
of the recording span as of the reference channel.
Use the [

] and [

] keys to select the maximum value

(SPAN R) in the same way as described under ‘SPAN L
setting
’. Note that SPAN L cannot be equal to SPAN R.
Press the [ENT] key.

[

] is displayed and the setting is completed.

Press the [ENT] key to set another channel; or press the
[DISP] key for three seconds to return to the OPERATION
mode.

NOTE

If you press the [ESC] key in the middle of setting, the indication returns to the

[

] and already set data will be aborted.

6.1 Setting Input Range and Recording Span (RANGE)

PRINT

DISP

ESC

3 sec

PRINT

PRINT

PRINT

DISP

ESC

3 sec

PRINT

FEED

PRINT

FEED

CH UP

CH UP

CH UP

CH UP

CH UP

CH UP

CH UP