Checks and Adjustments

1705A Spectrum Monitor

5- 13

ALIGN 2nd
MARKER HERE

DISREGARD

#1

GRATICULE

LINE

#2

GRATICULE

LINE

Figure 5-5: Aligning L- Band time markers with the graticule lines. The graticule
begins with 0 while the time markers are counted 1, 2, 3, etc. The first and last
divisions of sweep should be disregarded

11. Check L--Band SPAN/DIV and Readout

REQUIREMENTS:

FULL —

1 marker/Division

±1 minor Division

10 MHz —

1 marker/5 Divisions

±1 Division

1 MHz —

1 marker/5 Divisions

±1 Division

Readout Accuracy —

±20 MHz.

a.

Set CENTER FREQUENCY to 1400 MHz, and set SPAN/DIV to 10
MHz.

b.

Set the leveled sine wave generator Frequency to 50 MHz.

c.

Position the readout caret to center of screen horizontally.

d. CHECK

-- for 1 marker every 5 Divisions,

±1 Division (adjust

CENTER FREQUENCY, if necessary).

e.

Repeat parts a. through d. with CENTER FREQUENCY at 950 MHz
and 1700 MHz.

f.

Set CENTER FREQUENCY to 1400 MHz.

g.

Set SPAN/DIV to 1 MHz.