Checks and Adjustments

5- 20

1705A Spectrum Monitor

3.

Adjust Geometry, Focus, and Astigmatism

Display the DAC test pattern by completing parts a. through c.

a.

Press both the 1705A SPAN/DIV buttons simultaneously to display the
Main Menu.

b.

Press either SPAN/DIV button to move the cursor to TEST, then press
the INPUT button to display the TEST menu.

c.

Press either SPAN/DIV button to move the cursor to DAC, then press
the INPUT button to display the DAC test pattern.

d.

Set the front-panel FOCUS control so that it is approximately at the
center of its rotation.

e.

ADJUST

-- R45 (GEOM) for 35V

ᐔ1V at pin 1 of J3.

f.

ADJUST

-- R11 (CTR FOCUS) and R49 (ASTIG) for the most

clearly-defined DAC test pattern. See Figure 5-8.

Tek

REF

--10

--20

--30

--40

--50

--60

--70

L

O

G

+ f

Figure 5-8: DAC check waveforms used to check Focus, Astigmatism, Geometry,
and setting of the Trace Rotation (DAC test pattern may vary in size from instru-
ment to instrument)

4.

Adjust Trace Rotation

a.

ADJUST

-- the front-panel TRACE ROTATION potentiometer for a

level display, using a crt graticule line as a reference.