Maintenance

6- 22

1705A Spectrum Monitor

BESCHLEUNIGUNGSSPANNUNG KLEINER ALS

20kV

DIE IN DIESEM GERAT ENTSTEHENDE

RONTGENSTRAHLUNG IST AUSREICHEND ABGESCHIRMT

TO AVOID ELECTRICAL SHOCK, THE POWER

CORD PROTECTIVE GROUNDING CONDUCTOR

MUST BE CONNECTED TO EARTH GROUND.

SPECTRUM MONITOR

70 MHz

INPUT

L-BAND

INPUT

LNB POWER

LNB POWER
ON
OFF
+18VDC@250mA

Remove these five (5)
screws to remove the
rear panel.

Secured by removeable
nuts inside the panel.

ON

WARNING

0.70 MAX

50/60Hz

90 - 250V REPLACE FUSE ONLY

WITH

250V 2A F TYPE

333--3990--01

Figure 6-9: Screws securing the rear panel

1.

Remove the five rear screws. See Figure 6-9. Screws securing the power
receptical use removeable nuts as fasteners inside the rear panel.

2.

Unsolder the bnc and ground connections. (If 1700F10 Field Upgrade is
installed, unsolder leads from the battery connector.)

3.

Unsolder the two leads to the +18 V switch.

4.

Remove the hex nut on the L--BAND INPUT connector.

5.

Pull the rear panel free from the chassis, be careful not to pull the unsoldered
wires.

6.

To replace, reverse the procedure.

1.

Remove the blue multiwire connector from J154.

2.

Remove the two screws holding the board in place. See Figure 6-10 for
location.

Removal of the Rear Panel

Removal of Front- Panel

Assembly