Checks and Adjustments

1705A Spectrum Monitor

5- 21

b.

Exit the DAC test pattern by first pressing the INPUT button to display
the TEST menu.

c.

Press either SPAN/DIV button to move the cursor to EXIT, then press
the INPUT button twice to return to a normal display.

A3 MAIN BOARD

(Refer to Figure 5-7 for adjustment locations)

5.

Adjust On-Board Regulated Power Supplies

a.

Connect the voltmeter ground lead to the rear-panel LNB POWER
switch ground lug, and the active lead to the --11.8 V test point (W8).
See Figure 5-7.

b. ADJUST

-- R99 (--11.8 V ADJ) for --11.8 V (--11.78 to --11.82 V).

c.

Connect the voltmeter active lead to the +11.8 V test point (W9). See
Figure 5-7.

d. ADJUST

-- R111 (+11.8 V ADJ) for +11.8 V (+11.78 to +11.82 V).

6.

Adjust Horizontal Gain

a.

Connect a shorting strap between W11 and W12. See Figure 5-9.

W11

W12

SHORTING

STRAP

Figure 5-9: Location of the shorting strap used when adjusting sweep length

b. ADJUST

-- R168 (HORIZ GAIN) for a sweep length of 10 major

Divisions. Use the --60 horizontal graticule line as the reference.

c.

Remove the shorting strap.