Checks and Adjustments

5- 4

1705A Spectrum Monitor

6.

Check Resolution Filter

b. CHECK -- that the marker width 0.6 Divisions down from the top is
3 Divisions,

±1 Division.

e.

CHECK

-- that the marker width 0.6 Divisions down from the top of the

signal is

≤2 minor Divisions.

f.

CHECK

-- that the 10 kHz marker amplitude matches the highest point

of the 300 kHz marker amplitude within 1 minor Division.

7.

Check 70 MHz Gain and Flatness

e.

CHECK

-- for a marker amplitude change of 2 Divisions,

±2 minor

Divisions when switching the leveled sine wave generator Amplitude
Multiplier between .01 and 0.1 (a 20 dB change).

i.

CHECK

-- that the tip of the marker is on the crt center line,

±1 minor

Division (vertically).

8.

Check Video Filter

d. CHECK -- that the baseline noise amplitude drops approximately 50%
when the VIDEO FILTER is ON.

9.

Check Sweep Speed

b. CHECK -- for a solid trace with almost no flicker.

d. CHECK -- for approximately 3 to 4 sweeps per second.

Alternate Method

e.

CHECK

-- for a ramp duration of 20 ms

±10 ms.

h. CHECK -- for a ramp duration of 200 ms

±100 ms.

10.

Check L--Band Linearity

e.

CHECK

-- for 10 frequency markers.

f.

CHECK

-- that each marker is within 1 minor Division of a major

graticule line.

11.

Check L--Band SPAN/DIV and Readout

d. CHECK -- for 1 marker every 5 Divisions,

±1 Division.

i.

CHECK

-- for 1 marker every 5 Divisions,

±1 Division.

o. CHECK -- that the marker is intensified.

q. CHECK -- that the marker is on screen.

Advertising