Maintenance

6- 16

1705A Spectrum Monitor

Table 6- 5: High Volts Supply Fault Symptoms

Symptom

Procedure

Unable to focus crt using the front-
panel control

Focus Amplifier Check

Unable to adjust crt intensity using the
front-panel control

Z--Axis Amplifier Check
Grid Drive Check

No crt display

High Voltage Oscillator Check
CRT Voltage Check

Load the Low Volts Supply with the instrument, or with the 20

Ω resistor as

stated at the beginning of the Troubleshooting Procedure.

Focus Amplifier Check

Unsolder and lift one end of R24.

Power up the Power Supply.

Using the digital multimeter, measure the voltage between TP1 and the collector
of Q1. It should be approximately --140 V.

Reconnect the lifted end of R24.

Z-Axis Amplifier Check

Unsolder and lift one end of R8.

Power up the Power Supply.

Using the digital multimeter, measure the voltage between TP1 and the collector
of Q4. It should be approximately +10 V.

Short together the base and emitter of Q5. The collector of Q4 should be
approximately +100 V.

Reconnect the lifted end of R8.

Grid Drive Check

Turn off the Power Supply. Use the digital multimeter’s diode check to test
CR1, CR2, CR3, CR5, and CR6 for shorts.

Power up the Power Supply.

Using the digital multimeter, measure the voltage between TP1 and the cathode
of CR5. It should vary between approximately +75 and +200 V as R58 (CRT
Bias) is adjusted.