9.3.1 Checking /Adjusting the Plate Level

Adjustment is required if the axis of the plate level is not perpendicular to the vertical axis.

l

Check

1) Place the plate level parallel to a line running through the centers of two leveling screws, say, A and

B. Use these two leveling screws only and place the bubble in the center of the plate level.

2) Rotate the instrument 180° or 200g around the vertical axis and check bubble movement of the

plate level. If the bubble has been displaced, then proceed with the following adjustment.

l

Adjustment

1) Adjust the level adjustment capstan screw, with the accessory adjusting pin and return the bubble

towards the center of the plate level. Correct only one-half of the displacement by this method.

2) Correct the remaining amount of the bubble displacement with the leveling screws.
3) Rotate the instrument 180° or 200g around the vertical axis once more and check bubble

movement. If the bubble is still displaced, then repeat the adjustment.

9.3.2 Checking /Adjusting the Circular Level

Adjustment is required if the axis of the circular level is also not perpendicular to the vertical axis.

l

Check

1) Carefully level the instrument with the plate level only. If the bubble of the circular level is centered

properly, adjustment is not required. Otherwise, proceed with the following adjustment.

l

Adjustment

1) Shift the bubble to the center of the circular level, by adjusting three capstan adjustment screws on

the bottom surface of the circular level, with the accessory adjusting pin.

Leveling screw A

Plate level

Leveling screw B

Level adjustment
capstan screw

Half amount of displacement

Capstan adjustment
screws

Bottom of the base