Operation and Description of Instrument Function

Average Temperature/Water Bottom Sensor and Converter

44

Installation and Operations Manual

VH92 Error Dis. Sel.

Item type: select

Default: 0_OFF

Selection: 0_OFF, 1_ON

Type of VH88 "Short Error Value" and VH89 "Open Error Value" display selection.

0_OFF: These 2 error messages will not be transmitted to the connected host gauge.

This function automatically excludes defect element in average temperature calculation.

1_ON: Error message will be transmitted to the host gauge. As a result, VH88 and 89"s

numeric error code will be displayed on host gauge default screen and may transmit to

upper receiver as well.

VH94 Polling Address

Item type: read and write

Default: 2

Range: 1 ~ 15

Polling address for HART communication

VH95 Manufacture ID

Item type: read only

Default: 17

The manufacturer ID for Endress+Hauser instrumentation.

VH96 Software Version:

Item type: read only

A display of installed software version.

VH97 Hardware Version

Item type: read only

A display of installed hardware version.

VH98 Below Bottom

Item type: select

Default: 0_OFF

Selection: 0_OFF, 1_ON

A type of error display when liquid level drops below #1 element (Bottom Point). Error

code "29" is displayed on VH80 and VH91 when 0_ON is selected.

VH99 Device Type Code

Item type: read only

Device type will be displayed.

190: Temperature measurement function only.