This parameter can be set to ON or OFF.
Selecting ON automatically sets certain
parameters for type approved require-
ments. Before setting LFT ON, the bal-
ance must be calibrated. See Calibration
Menu for details. For sealing method,
refer to Type Approved Balance Sealing
section.
1.
Access the LFT submenu.
2.
The display will show the current sta-
tus (ON/OFF). If ON is selected,
Auto-Zero Tracking will be locked on
to 0.5d. and Lock Switch default be-
comes Setup.
3.
Press
to change the status.
4.
Press
to accept the status.
5.
When
is released, the display
will show LFT.
To access a submenu:
1.
Repeatedly press
until the de-
sired submenu is displayed.
2.
Press
to select the displayed
submenu.
NOTE
NOTE
NOTE
NOTE
NOTE: You must use END to store any
changes you make to the Setup menu.
The following sections describe each item
on the Setup menu in detail.