Error Codes (Cont.)

Error Codes (Cont.)

Error Codes (Cont.)

Error Codes (Cont.)

Error Codes (Cont.)

CheckSum Errors

CheckSum Errors

CheckSum Errors

CheckSum Errors

CheckSum Errors

9.0

Bad factory checksum. If error persists, have the balance serviced.

9.5

Bad factory calibration checksum. If error persists, have the balance serviced.

9.6

Bad mode checksum. Turn the balance off using the front panel controls. If the
error persists, have the balance serviced.

9.7

Invalid setup data checksum. Check Setup, User and Print menu (when RS232
is installed) settings. If possible, try to enter menus and exit using END to restore
menu settings. May be caused by a faulty component, or in rare cases, a severe
static charge. If error persists, balance must be serviced.

9.8

Hardware error causing invalid calibration data checksum. Balance may need
recalibration - particularly linearity calibration. If error persists, balance must be
serviced.

9.9

Invalid temperature compensation checksum. Balance will work with default
temperature compensation data, however, error will occur each time balance is
turned on. Have balance serviced.

SERVICE INFORMATION

SERVICE INFORMATION

SERVICE INFORMATION

SERVICE INFORMATION

SERVICE INFORMATION

If the Troubleshooting section does not resolve or describe your problem, you will need
to contact an authorized Ohaus Service Agent. For Service assistance in the United
States, please call Ohaus Corporation toll-free at (800) 526-0659. An Ohaus Product
Service Specialist will be available to help you.