4.9 Percent Weighing

Place the reference tem on the pan (ex. 300g) then press %.

“100.00%” wll be dsplayed.

The weght of any tem subsequently placed on the pan wll

be dsplayed as a percentage (ex. 600g wll be dsplayed as

200%) of the orgnally establshed reference weght.

5. SCALE SETTINGS

• Press then release

COUNT and M+ at the same tme to enter nto the user-

selectable scale settngs (Setup mode).

In Setup mode:

o Press to step through avalable settngs

o Press

Enter

to accept the dsplayed settng and proceed to the next

parameter

o Press

Exit

to proceed to the next parameter wthout savng any

changes

• Re-start the scale after changng settngs n the Setup mode.

The followng parameters are avalable (llustrated dsplays reflect default settngs):

NOTES:

• It s normal for small weght changes to be dsplayed as large decmal

ncrements n %.

• To ext Percent Mode, press

Exit

.

5.1 Scale Increment

Sets the dsplayed scale ncrement (also known as readablty

or graduaton). Selectable setup values wll be model

dependent and are equvalent to wthn 30,000 to 3,000 scale

dvsons.

5.2 Backlight

Sets the actvaton mode of the backlght.

The followng settngs are avalable: AU (Auto-on wth tems

greater than 9d placed on the pan or any key s pressed; turns

off after 5 seconds of nactvty), on, oFF.

5.3 Zero Tracking Range

Sets the range n whch the zero readng s mantaned.

The followng settngs are avalable: 0.5d, 1d, 2d, 4d (d =

scale dvson)

5.4 Filtering

Sets the level n whch the Stable ndcaton turns on; the hgher

the settng, the faster the stablzaton tme.

The followng settngs are avalable: 0, 1, 2, 3 (levels)