10
Calibration Procedure
Calibration Procedure
Calibration Procedure
Calibration Procedure
Calibration Procedure
Before beginning the procedure, make
sure that the calibration weight(s) is on
hand.
NOTE
NOTE
NOTE
NOTE
NOTE: DO NOT REMOVE CALIBRA-
TION WEIGHT UNTIL WEIGHT VALUE
IS DISPLAYED ON THE BALANCE.
Span Calibration
Span Calibration
Span Calibration
Span Calibration
Span Calibration
To start SPAN
SPAN
SPAN
SPAN
SPAN calibration, the BAL-
ANCE MUST BE ON
ON
ON
ON
ON.
1.
Press and hold
until CAL is
displayed, then release it.
2.
When
is released, C 0 will
be displayed indicating that no
weight should be on the platform.
3.
Press
. The display will
show -C- momentarily, followed by
the value of the weight (200g) which
must be placed on the platform.
4.
Place the required weight on the
platform and momentarily press
. The display will show -C-
while the balance recalibrates.
5.
When the weight on the platform is
displayed with the current unit indi-
cator, (g or oz), the balance is
recalibrated.