Linearity calibration – Ohaus C 305 ELECTRONIC BALANCE User Manual

Page 37

Advertising
background image

11

Linearity Calibration

Linearity Calibration

Linearity Calibration

Linearity Calibration

Linearity Calibration

To start LINEARITY

LINEARITY

LINEARITY

LINEARITY

LINEARITY calibration, the BAL-

ANCE MUST BE OFF

OFF

OFF

OFF

OFF.

NOTE

NOTE

NOTE

NOTE

NOTE: DO NOT REMOVE CALIBRA-
TION WEIGHT(S) UNTIL WEIGHT VAL-
UE IS DISPLAYED AND UNIT INDICA-
TOR APPEARS.

1.

Press and hold

, release

when CAL is displayed.

2.

When

is released, C 0 will

be displayed indicating that no

weight should be on the platform.

3.

Press

. The display will

show -C- momentarily, followed by

the value of the weight which must

be placed on the platform.

4.

Place the required weight on the

platform, then press

. The

display will show -C- momentarily,

followed by the next weight to be
placed on the platform.

NOTE: Display values
shown are for the Model
C305.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: DO NOT disturb the bal-
ance when -C- is displayed. Distur-
bances will result in improper cali-
bration.

5.

Place the required weight on the

platform, then press

. The

display will show -C- while the bal-
ance recalibrates.

6.

When the weight on the platform is
displayed along with the current unit
indicator (g or oz), the balance is

Advertising