Linearity calibration – Ohaus C 305 ELECTRONIC BALANCE User Manual
Page 37

11
Linearity Calibration
Linearity Calibration
Linearity Calibration
Linearity Calibration
Linearity Calibration
To start LINEARITY
LINEARITY
LINEARITY
LINEARITY
LINEARITY calibration, the BAL-
ANCE MUST BE OFF
OFF
OFF
OFF
OFF.
NOTE
NOTE
NOTE
NOTE
NOTE: DO NOT REMOVE CALIBRA-
TION WEIGHT(S) UNTIL WEIGHT VAL-
UE IS DISPLAYED AND UNIT INDICA-
TOR APPEARS.
1.
Press and hold
, release
when CAL is displayed.
2.
When
is released, C 0 will
be displayed indicating that no
weight should be on the platform.
3.
Press
. The display will
show -C- momentarily, followed by
the value of the weight which must
be placed on the platform.
4.
Place the required weight on the
platform, then press
. The
display will show -C- momentarily,
followed by the next weight to be
placed on the platform.
NOTE: Display values
shown are for the Model
C305.
NOTE:
NOTE:
NOTE:
NOTE:
NOTE: DO NOT disturb the bal-
ance when -C- is displayed. Distur-
bances will result in improper cali-
bration.
5.
Place the required weight on the
platform, then press
. The
display will show -C- while the bal-
ance recalibrates.
6.
When the weight on the platform is
displayed along with the current unit
indicator (g or oz), the balance is