Linearity calibration – Ohaus C 305 ELECTRONIC BALANCE User Manual

Page 56

Advertising
background image

11

Linearity Calibration

Linearity Calibration

Linearity Calibration

Linearity Calibration

Linearity Calibration

To start LINEARITY

LINEARITY

LINEARITY

LINEARITY

LINEARITY calibration, the

BALANCE MUST BE OFF

OFF

OFF

OFF

OFF.

NOTE

NOTE

NOTE

NOTE

NOTE: DO NOT REMOVE CALIBRA-
TION WEIGHT(S) UNTIL WEIGHT
VALUE IS DISPLAYED AND UNIT IN-
DICATOR APPEARS.

1.

Press and hold

, release

when CAL is displayed.

2.

When

is released, C 0

will be displayed indicating that no

weight should be on the platform.

3.

Press

. The display will

show -C- momentarily, followed by

the value of the weight (100g) which

must be placed on the platform.

4.

Place the required weight on the

platform, then press

. The

display will show -C- momentarily,
followed by the next weight (200g)
to be placed on the platform.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: DO NOT disturb the bal-
ance when -C- is displayed. Dis-
turbances will result in improper
calibration.

5.

Place the required weight on the

platform, then press

. The

display will show -C- while the bal-
ance recalibrates.

6.

When the weight on the platform is
displayed along with the current
unit indicator (g or dwt), the bal-
ance is recalibrated.