Overview
85
Electrodes
Socket acc. to DIN 19 262 or BNC
4-mm socket for reference electode
Temp. probe
4-mm socket for temperature probe
Interface/
analog output
Bidirectional RS 232 interface or ana-
log output with autom. recognition of
the PC, printer, or recorder connected
Plug-in power
supply
2 pole special Friwo
Connectors
AutoCal TEC
With Technical buffers, pH values
2.00; 4.00; 7.00; 10.00 at 25 °C;
buffers are temperature compensa-
ted in the range 0 ... 90 °C
AutoCal DIN
With standard buffers according to
DIN 19 266/NIST, pH values:
1.68; 4.01; 6.87; 9.18 at 25 °C;
buffers are temperature compensa-
ted in the range 0 ... 90 °C
ConCal
With any buffers; one and two-point
calibration, manual acceptance of
measured values and temp. input
Print principle
Matrix thermoprint
signs per line
26
printing width
46.1 mm
width of printing paper
58 mm
printing speed
40 signs/sec
Calibration modes
Printer