4 concal – Xylem CG 843P User Manual
Page 37

Operation
109
4.3.4 ConCal
Two-point
calibration
Use two buffer solutions for this procedure:
l
pH 7.0 ± 0.5
l
any other buffer solution
Note
Steps 2 and 9 are not required if you use a temperature
probe.
1 Press the <CAL> key repeatedly until
ASY
and the
ConCal
function display appear.
2 If necessary, set the temperature of the buffer solu-
tion by pressing <
▲
> <
▼
>.
3 Submerse the pH electrode in the first buffer solution
pH 7.0 ± 0.5.
4 Press the <run/enter> key.
The measured pH value appears on the display.
5 Set the nominal pH value of the buffer solution (at the
current temperature) by pressing the <
▲
> <
▼
> keys.
6 Press the <run/enter> key.
The value of the asymmetry (mV) and the sensor
symbol appear on the display.
7 Press the <run/enter> key.
SLO
(pe) appears on the display.