4 clearing the storage – Xylem CG 843P User Manual
Page 47

Operation
119
4.4.4 Clearing the storage
This function can erase the stored data records. 200 mem-
ory locations will then become available again.
Note
The
Clear store
function only appears if data records have
already been stored in the storage. Otherwise, the pH meter
automatically changes to the measuring mode.
In order to delete all the data records, proceed as follows:
Note
The calibration data remain in the storage and can be called
up via the calibration protocol.
1 Switch off the instrument.
2 Press and hold down the <STO> key.
3 Press the <on/off> key.
The
display test
appears briefly on the display.
4 Confirm the clearing process by pressing <run/
enter>.
Pressing any other key stops the clearing process
and the data records remain in the storage.