What to do if...
136
Buffer solutions:
– Incorrect buffer solutions
– Change calibration
procedure
– Buffer solutions too old
– Only use once.
Note the shelf life
– Buffer solutions depleted
– Change solutions
Cause
Remedy
pH electrode:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
Sample:
– pH value not stable
– Measure with air
excluded if necessary
– Temperature not stable
– Adjust temperature if
necessary
Electrode + sample:
– Conductivity too low
– Use suitable electrode
– Temperature too high
– Use suitable electrode
– Organic liquids
– Use suitable electrode
Cause
Remedy
– Calibrating interval
expired
– Newly calibrate measuring
system
No stable
measured value
Sensor symbol
flashes