Xylem CG 843P User Manual
Page 49

Operation
121
Setting the
data transmission
interval
The interval is set to OFF in the factory.
To start the data transmission, set up an interval (5 s, 10 s,
30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min):
Note
If the
AutoStore
function is active, the data transmission is
performed according to the setting of the storage interval
(Int1). Set the storage interval (Int 1) to OFF to activate the
data transmission
interval (Int 2).
Note
You can also set the
data transmission
interval (Int 2) in the
Configuration
menu (see page 124).
1 Press and hold down the <run/enter> key.
2 Press the <auto read> key.
Int.2
appears on the display.
3 Press <
▲
> <
▼
> to set up the required time interval
between storage processes.
4 Confirm with <run/enter>.
The instrument changes automatically to the pH
measuring mode.