Xylem CG 843P User Manual
Page 29

Operation
101
Calibration
evaluation
After the calibration, the instrument automatically evaluates
the current status. The asymmetry and slope are separately
evaluated. The worst evaluation appears on the display.
Display
Asymmetry
[mV]
Slope
[mV/pH]
-15 ... +15
58 ... 60.5
-20 ... +20
57 ... 58
-25 ... +25
56 ... 57
or
60.5 ... 61
Clean the electrode accord-
ing to the electrode operat-
ing manual
-30 ... +30
56 ... 50
or
61 ... 62
Clear the fault according to
chapter 6 W
HAT
TO
DO
IF
...
< -30 or
> 30
< 50 or
> 62
1 Switch on the instrument by pressing <on/off>.
2 Connect the pH electrode to the instrument.
3 Keep the buffer solutions ready.
4 Adjust the temperature of the solutions and measure
the current temperature if the measurement is per-
formed without a temperature probe.
Preparatory
activities