Operation

HandyLab 200

30

ba77063e01

03/2014

Measurement /
Measuring cell /
Type

Measuring cell used

Cal

Measuring cells the cell
constant of which is deter-
mined by calibration in the
KCL control standard solu-
tion.
Calibration ranges:
0.450 to 0.500 cm

-1

and

0.800 to 0.880 cm

-1

The currently valid cell con-
stant is displayed in the sta-
tus line.

LF313T

Measuring cell LF313T,
nominal cell constant
0.100 cm

-1

.

The cell constant can be
adjusted in the range from
0.090 to 0.110 cm

-1

.

LF213T

Measuring cell LF213T,
nominal cell constant
0.010 cm

-1

.

The cell constant is perma-
nently set.

man

Any measuring cells with
freely adjustable cell con-
stants in the range from
0.250 to 25.000 cm

-1

.

Measurement /
Measuring cell /
Cell const. man

0.250 to
25.000 cm

-1

Display and setting option
of the cell constant of any
measuring cells (man).

Measurement /
Measuring cell /
Cell const. LF313T

0.090 to
0.110 cm

-1

Display and setting options
for the cell constant of the
measuring cell LF313T.

Measurement /
Temp. comp. (TC) /
Method

nLF
Lin
Off

Procedure for temperature
compensation (see section
4.4.3).

This setting is only avail-
able for the measured
parameters,

ϰ

and

ρ

.

Menu item

Possible set-
ting

Explanation