6 displaying calibration records, Displaying calibration records – Xylem HandyLab 200 User Manual
Page 33
Advertising

HandyLab 200
Operation
33
ba77063e01
03/2014
4.4.6
Displaying calibration records
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the <CAL
_
> key.
4
Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
5
Start the measurement with <ENTER>.
The [AR] status indicator is displayed. The measured parame-
ter flashes.
6
Wait for the end of the AutoRead measurement or accept the
calibration value with <ENTER>.
The calibration record is displayed.
7
Using <F1>/[Continue] or <ENTER>, switch to the measured
value display.
ϰ
0.479 1
/cm
25.1
°C
15.03.2014
10:00