6 reset, 1 resetting the measurement settings, Reset 4.6.1 – Xylem HandyLab 200 User Manual

Page 38: Resetting the measurement settings

Advertising
background image

Operation

HandyLab 200

38

ba77063e01

03/2014

4.6

Reset

You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.

4.6.1

Resetting the measurement settings

Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.

The following settings for conductivity measurements are reset to the
default settings with the Reset function:

The sensor settings are reset under the Reset menu item in the menu
for calibration and measurement settings. To open the settings, acti-
vate the relevant measuring window in the measured value display and
press the <F1>[Menu] key shortly.

Setting

Default settings

Cal. interval

150 d

Measured parameter

ϰ

Cell constant (c)

calibrated: 0.475 cm

-1

adjusted: 0.475 cm

-1

calibrated: cm

-1

adjusted:

Temperature compensation

nLF

Reference temperature

25 °C

Temperature coefficient (TC) of
the linear temperature compen-
sation

2.000 %/K

TDS factor

1.00

Stability control

On

Temperature unit

°C