Xylem HandyLab 200 User Manual
Page 31

HandyLab 200
Operation
31
ba77063e01
03/2014
Measurement /
Temp. comp. (TC) /
Linear coeff.
0.000 ...
3.000 %/K
Coefficient of the linear
temperature compensa-
tion.
This setting is only avail-
able when the linear tem-
perature compensation is
set.
Measurement /
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only avail-
able for the measured
parameters,
ϰ
and
ρ
.
Measurement / TDS
factor
0.40 ... 1.00
Factor for TDS value
Measurement /
Stability control
On / Off
Switches on or off the auto-
matic stability control dur-
ing measurement (see
section 4.3.3)
Measurement /
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperature values are
displayed with the selected
unit.
Reset
-
Resets all sensor settings
to the delivery condition
(see section 4.6.1).
Menu item
Possible set-
ting
Explanation