Configure the 3100, Cell constant, 3 configure the 3100 – YSI 3100 User Manual

Page 10

Advertising
background image

6

16.34

µ

S

24.8

°C

3.3 Configure the 3100

Before operating the 3100, or whenever you change cells, you must configure the 3100 to match
the cell used. You must enter the manufacturer’s stated (or your manually calculated) cell
constant (K) as shown below (Cell Constant).
The default configuration is as follows:

Cell constant of K = 1

Temperature compensation corrected to 25°C using a coefficient of 1.91%/°C.

If you are using a cell with a cell constant of K=1 (such as the YSI 3252, 3253, 3254, 3401,
3403, 3417, or 3445), the cell constant is already set correctly. If, however, you are using a cell
with a different cell constant (such as the YSI 3256, 3255, 3402, 3418, 3440, or 3446), you must
enter the manufacturer’s stated cell constant for the cell that you are using (K= 0.1, 10.0 etc.) as
shown below (Cell Constant).

Cell Constant

Follow these steps to change the cell constant.

1.

With the instrument on, press and release the [DOWN ARROW] and [MODE] keys at the
same time. The CAL symbol will appear at the bottom left of the display and the large portion
of the display will show

1.91

% (or a value set previously using Advanced Setup).

1.91

%

22.7

°C

CAL

2.

Press and release the [MODE] key. The large portion of the display will show

25.0C

(or a

value set previously using Advanced Setup).

3.

Press and release the [MODE] key again. The large portion of the display will show

1.00

(or a

value set previously using Advanced Setup).

Temperature: °C

Conductivity

Cal symbol